Skip to main content

Biosafety Cabinets Class II A2

BIO1+, BIO2+, BIO3+ & BIO4+

Cruma BIO+ biosafety cabinets for microbiological safety and environmental protection are certified by Tüv Nord pursuant to Standard EN 12469:2000. These are CLASS II TYPE A2 biosafety cabinets equipped with UV light as standard and which were developed to protect the personnel and the environment when material from risk groups 1, 2 and 3 is being handled and, at the same time, to protect the material in question from external contamination.

Features

Yes, they all fit and are standard.

  • Absolute filters H14 with efficiency 99,995 in particles of 0,3 microns.
  • Recirculation flow balanced with the discharge flow, to guarantee 25 changes / minute of sterile air in the working space.
  • Average airspeed in the façade, at a height of 195 mm, equal to or greater than 0.5 m / sec.
  • Containment factor greater than 1.5 x 105
  • Degree of cleaning DC according to EN 12469.
  • Illumination of the working plane greater than 900 lux.
  • Sound pressure level according to ISO 1102: less than or equal to 58 dB (A).
  • Power supply standard 230 V 50 Hz.
  • Current admitted for each service outlet 3A.
  • Analog microprocessor watch-dog.
  • CE Certification according to Directive in Machines 89-392
  • Certified EN 12469-2000
Membrane keypad: opening and closing controls of the front glass, activation of service ports, electrical outlet, electro gas valve, lighting and key ready to activate the UV lamp “Cruma
Emergency Stop key
Electrical inlet
Vacuum tap
Gas tap
Optional stand
UV lamp

Uses

But we’re sure you’ll find more.

  • Manipulation of microorganisms, bacteria, fungi, viruses
    and parasites Risk categories 1, 2 and 3.
  • Isolation and sample culture
  • Quantification methods
  • Microscopy techniques and sample preparation – Identification and classification of microorganisms
  • Genetic Manipulation

Alarms and warnings

Alarms. Multilevel alarms, with redundancy functions
Alarms for low airflow and wrong front window position
Permanent  display of working conditions

Functional Specifications

Continuous monitoring of safety parameters.
Continuous monitoring of front barrier air flow for the highest operator safety
Maximum ventilation stability also in the presence of temporary turbulence and / or progressive variations due to saturation of the filters.
Electronic fan control based on a volumetric sensor of the air flows whose signal is processed by a microprocessor.
Software predisposed for the management of the semi-automatic cycle decontamination of the cabin.

Simple and effective

Sloped front design for the highest operational comfort. Sloped back side of the working chamber for the best down flow disbution (cabinet carcass EN12298 tested and certified for air tightness).
Maintenance and change of filters with frontal access.
Utilities inlets from the top of the cabinet

I Want More Information and Prices

    Accept the privacy policy





    *Required fields

    Work fully protected

    No need to worry. It’s totally safe.

    When in operation a downward stream of sterile air (filtered through the first H14 HEPA filter) sweeps over the stainless work surface laminarily. On its way down this stream creates a protective barrier of air towards the inside of the cabinet which prevents the particles and/or organisms from being released to the outside, protecting the operator from possible contamination.

    Both streams of air are pushed though grids located in the work surface over to a plenum located in the upper part of the cabinet where the air is distributed. 30% is released to the outside through a second HEPA filter and the remaining 70% is recirculated once again towards the inside of the cabinet through the first filter.

    Models and Technical Specifications

    Because size does matter. And safety even more.

    Model BIO1+ BIO2+ BIO3+ BIO4+
    Internal Sizes (width/depth/high) 925x580x700 1230x580x700 1530x580x700 1840x580x700
    External Sizes (width/depth/high) 1075x795x1450 1380x795x1450 1685x795x1450 1990x795x1450
    Filtration columns 1 1 1 1
    Filter number 2 2 2 2
    Fans 1 1 1 1
    Average volume of treated air/hour +/-350 m3/h +/-400 m3/h +/-600 m3/h +/-600 m3/h
    Average face velocity 0.50 m/s 0.50 m/s 0.50 m/s 0.50 m/s
    Total electrical power consumption 260W 360 W 500 W 650 W
    Voltage-Frequency 110-220 V / 50-60 Hz 110-220 V / 50-60 Hz 110-220 V / 50-60 Hz 110-220 V / 50-60 Hz
    Light intensity Fluorescent +900 Lux/36 W Fluorescent +900 Lux/36 W Fluorescent +900 Lux/36 W Fluorescent +900 Lux/36 W
    UV Lamp Serial Serial Serial Serial
    Noise level <49 dB <50 dB <56 dB <58 dB
    Weight 227 Kg 256 Kg 310 Kg 390 Kg
    Packaging volume 2 m3 2.39 m3 2.5 m3 3.47 m3
    Warranty years 2 2 2 2
    Download Technical Specs PDF TECHNICAL DATA SHEET BIO1+ TECHNICAL DATA SHEET BIO2+ TECHNICAL DATA SHEET BIO3+ TECHNICAL DATA SHEET BIO4+
    OFFER REQUEST I WANT MORE INFO & PRICES I WANT MORE INFO & PRICES I WANT MORE INFO & PRICES I WANT MORE INFO & PRICES

    I Want More Information and Prices

      Accept the privacy policy





      *Required fields

      Serial Equipment

      All you need to work in safe conditions.

      Equipment
      Counter
      Auxiliary electrical service socket x2 units
      Combustibile gas solenoid valve
      Pre-installation for combustible gas and vacuum line (taps not included)
      UVC lamp (backwall mounted)

      Optional Equipment

      Details that matter

      Optional Equipment
      Gaseous fluids taps.
      Passive extraction kit for external expulsion.
      Motorized Kit for external extraction expulsion.
      Kit motorized bike extraction fan remote for external expulsion.
      Activated carbon filters or additional HEPA expulsion.
      Tubular table support.

      HEPA Filters

      Double-H14 HEPA filter to ensure the air remains sterile and health is unaffected

      * To view the technical sheets of our filters, please contact Cruma sales department at export@cruma.es
      – It is the user’s responsibility to correctly manage the waste generated from filters based on the use given to them and according to the applicable requirements.
      – The supply of spare parts is guaranteed during a period of 10 years

      I want more information and prices