To ensure the reliability of the tests, it is essential that the work carried out inside the cabinet does not become contaminated. To create an environment that is safe and particle-free, the vertical laminar flow cabinets we manufacture produce a stream of Class ISO 5 sterile air. This air passes through a H14 HEPA filter (manufactured pursuant to Standard EN-1822) at a constant speed, travelling downward through the cabinet in a vertical laminar flow and producing sterile air, protecting the product from external contamination and possible cross-contamination between samples.
The cabinets that are equipped with our filtration system, which has a minimum efficiency rate of 99.995% for 0.3 µm particles in the upper part of the filter and an extraction rate of 100% as regards the air flow to the outside.
Sample Protection. Total Sterility
Our vertical laminar flow cabinets guarantee a laminar stream of air with class ISO 5 air purity pursuant to Standard EN14644-1 thanks to our intelligent H14 HEPA filters, providing a minimum efficiency rate of 99.995% for 0.3 µm particles.
Flexible
These can be used in areas where it is difficult to extract contaminated air, such as the lower floors of buildings, and they can be placed in different locations, which it means that they can be easily relocated if needed.
Economic Savings and Easy to Set-Up
No expensive works are necessary to install the cabinet which means there are no additional costs and no need to interrupt a laboratory working at full capacity.
Energy Savings
Unlike other brands, our cabinets do not consume the air conditioning of the laboratory, which substantially reduces energy costs: the filtered air is not expelled to the outside, but recirculated back into the laboratory free of contaminating particles.
Features
Yes, they all fit and are standard.
Manufacture: the external surfaces are made of cold-rolled steel, coated with white epoxy, enamelled in ovens at 200ºC and resistant to the chemical compounds used for sterilization.
Work Surface: the surface of the work chamber is made of white secured glass, highly resistant to impacts and chemical attacks.
Absolute H14 HEPA Filter with a flow equalizer and an efficiency rate of over 99.995% for particles measuring 0.3 microns. Standards MIL-F51068C and BS 3298.
Fan motors: double aspiration centrifugal fans, capable of providing a laminar flow in the work space, above all as regards dealing with filter saturation to a maximum value of 400 Pascal.
Control Panel: Soft-touch keyboard with a switch for the LED lamps, UV ultraviolet lights, a maintenance system to control the airspeed through the HEPA filter that is electronically self-regulated and a control key to reduce speed in ECO mode.
Lighting: LED lamps located in the centre with a prismatic diffusing screen to improve the distribution of light on the work surface.
Uses
But we’re sure you’ll find more.
- Laboratory food industries in general
- Transfer of drugs in the pharmacy department
- Hematology and clinical analysis
- Filled with antibiotics and injectable drugs (except cytostatic) microscopic analysis
- Tissue culture bags reconstitution of parenteral methods of assisted fertilization techniques
- Micropropagation, Plant cell cultures, Water quality control, manufacturing of electronic devices…
127x34mm LCD Display
Type of filter installed
Working hours
Wxpiration date and next revision date
Countdown timer
Clock and Calendar
New features and components
Initial cycle flow adequacy and final purge cycle
Fault LED
Control of air flow through Microprocessor
HEPA filters with electronic chip
LED illumination
New alarms and scheduled warnings
Next validation warning
Few hours of filter life warning
Countdown timer warning
Expired filter alarm by hours & by date
Equipment without filter alarm
Clean Area
Protect your samples without cross contamination.
In the vertical laminar flow cabinets the stream of air passes through the H14 HEPA filter (manufactured pursuant to EN-1822) at a constant speed, travelling vertically through the cabinet in a laminar flow and producing a Class ISO 5 sterile environment (former American Federal Standard 209E) and protecting the product from external contamination and possible cross-contamination between samples.
Single-stage filtration system with a minimum efficiency rate of 99.995% for particles of 0.3 µm in the upper part with an extraction rate of 100% as regards the air flow to the outside. Total protection for the sample.
Models and Technical Specifications
All sizes for all needs.
Model | 670FL | 870FL | FL-1 | FL-2 |
Internal Sizes (Width/Depth/High) | 575x560x630 | 775x560x740 | 710x556x610 | 1110x556x610 |
Medidas externas (Width/Depth/High) | 600x600x930 | 800x600x1125 | 800x600x1137 | 1200x600x1137 |
Filter columns | – | – | – | – |
Filter number | 1 | 1 | 1 | 1 |
Fans | 1 | 1 | 1 | 1 |
Average volume of treated air/hour | – | – | – | – |
Average face velocity | 0.35 m/s | 0.35 m/s | 0.40 m/s | 0.40 m/s |
Internal volume of the cabinet | 0.181m3 | 0.287m3 | 0.236m3 | 0.376m3 |
Renewals inside the cabinet/min. | – | – | – | – |
Total electrical power consumption | 55 W | 55 W | 226 W | 226 W |
Voltage-Frequency | 110-220 V / 50-60 Hz | 110-220 V / 50-60 Hz | 110-220 V / 50-60 Hz | 110-220 V / 50-60 Hz |
LED light intensity | 1300 lm 15 W | 1300 lm 15 W | 900 Lux 16 W | 900 Lux 16 W |
UV Lamp | 1x15W | 1x15W | 1x15W | 1x15W |
Sound level | 57 ±3 dB | 50 ±3 dB | 50 dB | 50 dB |
Weight | 65 Kg | 92 Kg | 112 Kg | 142 Kg |
Packaging volume | 0.28 m3 | 0.42 m3 | 0.74 m3 | 0.95 m3 |
Download Technical Specs PDF | TECHNICAL DATA SHEET 670FL | TECHNICAL DATA SHEET 870FL | TECHNICAL DATA SHEET FL-1 | TECHNICAL DATA SHEET FL-2 |
OFFER REQUEST | I WANT MORE INFO & PRICES | I WANT MORE INFO & PRICES | I WANT MORE INFO & PRICES | I WANT MORE INFO & PRICES |
Serial Equipment
Full Equip Cabinets
Equipment | Description |
---|---|
Electronic circuit with large format LCD display | Security levels: level 1 for users and level 2 for maintenance users |
Filters controlled by microprocessor | The filters are provided with a microchip with mini USB connection identifying the type of filter installed, the expiry date and the serial number |
Illumination | 96 LED high light intensity and low consumption 16 Watts / 900 Lux |
Digital timer with audible warning | Hourmeter for the control and warning of the work in the cabinet |
Clock and Calendar | Display of date and time |
Work surface 1 | Liquid retention tray (2-10 liters) with working surface in white tempered glass |
HEPA H-14 filter | Manufactured according to EN 1822 with a minimum efficiency of 99.998% for particles of 0.3 microns |
G4 Prefilter | Pre-filtering class G4 biofibre synthetic blanket |
Warranty | 7 years |
Optional
Details that matter.
Equipment | Description |
---|---|
Movilair | Stand with wheels and internal tray in Epoxy coated steel |
Tubular steel stand | Support stand in Epoxy coated steel |
Working surface 2 | Spill retention tray (2-10 liters) with working surface in phenolic resin |
Working surface 3 | Spill retention tray (2-10 liters) with working surface in inox steel |
Transparent rear back pannel | Transparent polymethylmethacrylate rear pannel 8 mm thick (light transmission of 93%). Ideal for teaching sessions. |
Voltage / Frequency | 125 V / 50 Hz |
Filter test kit | Dräger pump with reactive colorimetric tubes (pack 10u) |
Cruma Filters
HEPA filter H-14 for a pure air. Pristine air.
Absolute H14 HEPA Filter with a flow equalizer and an efficiency rate of over 99.995% for particles measuring 0.3 microns. Standards MIL-F51068C and BS 3298. With an electronic chip to provide both the operators and the samples with the greatest possible degree of safety.
* To view the technical sheets of our filters, please contact Cruma sales department at export@cruma.es
– It is the user’s responsibility to correctly manage the waste generated from filters based on the use given to them and according to the applicable requirements.
– The supply of spare parts is guaranteed during a period of 10 years